Enmet MEDAIR 2200 User Manual Page 31

  • Download
  • Add to my manuals
  • Print
  • Page
    / 40
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 30
M
ED
A
IR
2200
ENMET Corporation
28
6.0 Technical Data and Specifications
Electrical Power
15 Amp fused branch circuit
100-240 V
AC
0.9 A
50/60 Hz
Board Mounted Fuse FH2, 0.630A, 5 x 20mm
Storage and Transport
Temperature:
-20° to +60°C (-4° to +140°F)
preferred
0° to +20°C (32° to 68°F)
Relative Humidity
0 - 99% RH, non-condensing
Atmospheric Pressure
20 to 36 inHg (68 to 133 kPa)
Operation
Temperature:
0° to +40°C (32° to +104°F)
Relative Humidity
0 - 99% RH, non-condensing
Atmospheric Pressure
20 to 36 inHg (68 to 133 kPa)
Air Line Pressure
55
PSI
(± 5
PSI
)
Mechanical
Dimensions:
11 x 9 x 6 inches (4.3 x 3.5 x 2.4 cm)
Weight:
8 lbs (3.6 kg)
Material:
Engineered thermoplastic with hinged front cover
Strain relief: 5-12 mm OD
Outputs
Relays:
SPDT
Resistive Load Inductive Load
10A at 110 VAC 7.5A at 110 VAC
10A at 30 VDC 5A at 30 VDC
Analog: 4-20 mA x 3
Digital:
RS-232 – Modbus
RS-485 – Modbus
Audio: 95 db at 2 ft
Sensors Type Range Response Time Life
CO 0 – 50 ppm T
90
= 30 seconds 1 – 3 years
Dew Point -112 - +68°F T
90
= 10 seconds for –40°F to 50°F step change 2 years
O
2
0 – 30% T
90
= 15 seconds 1 – 2 years
CO
2
0 – 5000 ppm T
90
= 30 seconds 3 – 5 years
HC 0 – 100 ppm T
90
= 30 seconds 1 – 2 years
N
OTE
: All specifications stated in this manual may change without notice.
Page view 30