ENMET Corporation Spectrum ON-LINE64.0 Maintenance4.1 Maintenance MenuFrom the operational display, press the OPTION pushbutton four times; "go S
Spectrum ON-LINE ENMET Corporation74.1.1 KeyEntrance to the maintenance menu is guarded with a four digit numerical key. The factory default setting
ENMET Corporation Spectrum ON-LINE8Sensor, Bottom View4.1.3 CalibrationNOTE: Calibration must be performed at normal room temperature (20-25°C) for op
Spectrum ON-LINE ENMET Corporation94.1.4 Changing the Alarm LevelA valid key entry sets the instrument at the "run ZERO" location of the mai
ENMET Corporation Spectrum ON-LINE105.0 Replacement Part NumbersENMET part numbers for replacement parts and accessories:Calibration EquipmentCylinder
Spectrum ON-LINE ENMET Corporation116.0 WARRANTYENMET warrants new instruments to be free from defects in workmanship and material under normal use fo
ENMET Corporation Spectrum ON-LINE12Appendix A: Calibration Data TablesTable 1: Gas Ranges, Alarm Points and Sensor LifeGas RangeAlarmlower limitAlarm
Spectrum ON-LINE ENMET Corporation13Table 2: Spectrum Calibration Voltage and Countdown TimesNOTE: The following are values used for production calibr
ENMET Corporation Spectrum ON-LINE14Appendix B: Interference GasesInterference Table for Selected Spectrum SensorsSPECTRUM Series for AMMONIA (3E-100
Spectrum ON-LINE ENMET Corporation15SPECTRUM Series for CHLORINE (3E sensor)The following gases cause a reading:InterferantGas/VaporInterferant Concen
ENMET Corporation Spectrum ON-LINE16SPECTRUM Series for Hydrogen Chloride (3E 30 sensor)The following gases cause a reading:InterferantGas/VaporInterf
Spectrum ON-LINE ENMET Corporation17SPECTRUM Series for Fluorine (3E 10 sensor)The following gases cause a reading:InterferantGas/VaporInterferant Con
ENMET Corporation Spectrum ON-LINE18SPECTRUM Series for Nitric Oxide (3E 100 sensor)The following gases or vapors have been tested and do not cause a
Spectrum ON-LINE ENMET Corporation19SPECTRUM Series for Ozone (3E 1 sensor)The following gases cause a reading:InterferantGas/VaporInterferant Concent
ENMET Corporation Spectrum ON-LINE20SPECTRUM Series for Sulphur Dioxide(4S sensor)The following gases cause a reading:InterferantGas/VaporInterferant
Spectrum ON-LINE ENMET Corporation21SPECTRUM Series for Phosgene (3E sensor)The following gases cause a reading:InterferantGas/VaporInterferant Concen
ENMET Corporation Spectrum ON-LINE22Notes:
PO Box 979680 Fairfield CourtAnn Arbor, Michigan 48106-0979734.761.1270 Fax 734.761.3220Returning an Instrument for RepairENMET instruments may be ret
Repair Return FormMailing Address:ENMET CorporationPO Box 979Ann Arbor, Michigan 48106Phone Number: 734.761.1270FAX Number: 734.761.3220Shipping Addr
Table of ContentsINTRODUCTION...
Spectrum ON-LINE ENMET Corporation1IntroductionThe SPECTRUM ON LINE is a small powered by an AC adapter operated single channel gas detection instrume
ENMET Corporation Spectrum ON-LINE21.2 Turn the instrument ONTurn the instrument ON, by plugging the power supply into a standard 110 VAC wall outlet.
Spectrum ON-LINE ENMET Corporation3Figure 1: SPECTRUM ON-LINE Features2.2 Installation2.2.1 Mount EnclosureMount enclosure at an appropriate location
ENMET Corporation Spectrum ON-LINE43.0 Operation3.1 Operational MenuThe operational menu diagram is shown in Figure 2. This menu is accessed with the
Spectrum ON-LINE ENMET Corporation53.2 Gas Concentration Display and AlarmsThe LCD furnishes a numerical display of the target gas concentration from
Comments to this Manuals